Advia Chemistry Systems (2013-11-11) Nov 11, 2013 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type III
Lot or serial number Companies Brand name Model or catalog number More than 10 numbers, contact manufacturer Manufacturer Siemens Healthcare Diagnostics Inc. 511 Benedict Ave. Tarrytown 10591 New York UNITED STATES D. Advia Chemistry System - Advia 2400 Analyzer 01714935 CA12240082 Manufacturer Siemens Healthcare Diagnostics Inc. 511 Benedict Ave. Tarrytown 10591 New York UNITED STATES B. Advia 1200 Chemistry System - Instrument 02475632 More than 10 numbers, contact manufacturer Manufacturer Siemens Healthcare Diagnostics Inc. 511 Benedict Ave. Tarrytown 10591 New York UNITED STATES C. Advia Chemistry System - Advia 1800 Analyzer 01412173 CA175542364 Manufacturer Siemens Healthcare Diagnostics Inc. 511 Benedict Ave. Tarrytown 10591 New York UNITED STATES A. Advia Chemistry System - Advia 1650 Analyzer 06363413
Recalled products
Advia Chemistry System - Advia 1650 Analyzer
Advia 1200 Chemistry System - Instrument
Advia Chemistry System - Advia 1800 Analyzer
Advia Chemistry System - Advia 2400 Analyzer
Siemens confirms that three issues were identified where the system may pass calibration and QC, when it should not. These include auto-calibration with flagged system related errors, auto-calibration with a lot change and reagent pack switching mid-calibration or mid-QC. Users are being advised when any of the three situations occur, to perform the calibration or QC again.