Health Products

Advia Chemistry Systems (2013-11-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
D. Advia Chemistry System - Advia 2400 Analyzer01714935
CA12240082
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
B. Advia 1200 Chemistry System - Instrument02475632
More than 10 numbers, contact manufacturer
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
C. Advia Chemistry System - Advia 1800 Analyzer01412173
CA175542364
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
A. Advia Chemistry System - Advia 1650 Analyzer06363413

Introduction

Recalled products
  1. Advia Chemistry System - Advia 1650 Analyzer
  2. Advia 1200 Chemistry System - Instrument
  3. Advia Chemistry System - Advia 1800 Analyzer
  4. Advia Chemistry System - Advia 2400 Analyzer

Reason

Siemens confirms that three issues were identified where the system may pass calibration and QC, when it should not. These include auto-calibration with flagged system related errors, auto-calibration with a lot change and reagent pack switching mid-calibration or mid-QC. Users are being advised when any of the three situations occur, to perform the calibration or QC again.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime