Health Products

Osseocare Pro (2013-10-24)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lotsA. Osseocare Pro CA 20:1 L MS KM
Manufacturer
Bien-Air Dental SA
Langgasse 60
Bienne
2504
SWITZERLAND
1700471

Introduction

Recalled products

A. Osseocare Pro CA 20:1 L MS KM

Reason

When using the Osseocare Pro Ipad application release 1.2.0.7, the grey button located on the foot pedal has no effect on the rotation direction. Pressing the grey button normally sets the motor in reverse rotation mode. However, when pressing the grey button with the software release 1.2.0.7, this does not change the rotation direction. This malfunction may result in the screw tap and/or implant being placed deeper than intended, which in the worst case could damage vital anatomical structures (Mandibular nerve and blood vesseles, or penetration in to the Lumen of maxillary sinus or nasal cavity).
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime