Health Products

Immulite and Immulite 1000 Systems (2013-11-07)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
335
Manufacturer
Siemens Healthcare Diagnostics Products Ltd.
Glyn Rhonwy
Llanberis, Gwynedd
LL55 4EL
UNITED KINGDOM
B. Immulite 1000 System - Folic Acid AssayLKFO1
335
Manufacturer
Siemens Healthcare Diagnostics Products Ltd.
Glyn Rhonwy
Llanberis, Gwynedd
LL55 4EL
UNITED KINGDOM
A. Immulite System - Folic Acid AssayLKFO1

Introduction

Recalled products
  1. Immulite System - Folic Acid Assay
  2. Immulite 1000 System - Folic Acid Assay

Reason

Siemens Healthcare Diagnostics identified the potential for an accelerated decline in counts per second for the Immulite/Immulite 1000 folic acid (LKFO1) kit lots 335, 336, and 337. This may lead to a failed adjustment before the expiration date, which is indicated on the kit label, is reached.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime