Health Products

Artis Zee Systems (2013-11-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
147097
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
B. Artis Zee Ceiling10094137
153854
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
C. Artis Zee Biplane10094141
137167
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
A. Artis Zee Floor10094135

Introduction

Recalled products
  1. Artis Zee Floor
  2. Artis Zee Ceiling
  3. Artis Zee Biplane

Reason

If the system is not turned off manually at regular intervals, an automatic reset of the large display controller can occur. During the reset, the system switches for approximately 1-2 minutes into the bypass fluoro mode. During this time, only the native fluoroscopic image is displayed.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime