Health Products

Fibrintimer (2013-11-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Siemens Healthcare Diagnostics Products GmbH
Emil-Von-Behring-Str. 76
Marburg
35041
GERMANY
A. Fibrintimer OVKF03

Introduction

Recalled products

A. Fibrintimer 

Reason

Incorrect standard human plasma (SHP) predilution in the title line of the standard calibration curve table in the application sheets of all endogenous coagulation factors were detected in reference guide versions 3.00, 3.01, 3.02. The stated predilutions do not fit to the recommended dilution set of the last calibration curve point.