Health Products

Artas Disposable Kit (2013-10-14)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
12490
Manufacturer
Restoration Robotics
128 Baytech Drive
San Jose
95134
California
UNITED STATES
A. Artas Disposable Kit FP-45384

Introduction

Recalled products

A. Artas Disposable Kit 

Reason

There is a discoloration on the tensioner pins that was observed prior to performing the Artas procedure.  This discoloration has been identified as a compound of iron and chlorine and found to be a remnant of the cleaning process used on the stainless steel (SS303) tensioner pins by the pin manufacturer.  Action has been taken to ensure that replacement product uses pins that have been properly cleaned and all potentially affected product in the field are to be returned and replaced.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime