C. Umbilical Line Draw Plus Arterial Blood Sampling Kit
4040-2
More than 1000 numbers, contact manufacturer
Manufacturer
Smiths Medical ASD Inc.
10 Bowman Drive
Keene
03431
New Hampshire
UNITED STATES
A. Pro-Vent Plus Arterial Blood Sampling Kit
4042LH
Introduction
Recalled products
Pro-Vent Plus Arterial Blood Sampling Kit
Dry Line Draw Plus Arterial Blood Sampling Kit
Umbilical Line Draw Plus Arterial Blood Sampling Kit
Reason
Smith Medical has identified an issue with ABL syringes when used with Radiometer brand blood gas analyzers. In August 2013, Smiths Medical began receiving a small number of complaints regarding use of the ABL syringes with Radiometer blood analyzers. The plunger tip in the ABL syringe does not remain stationary when the probe of the analyzer extends into the syringe. This allows air to leak into the sample at the luer, stopping the blood from entering the analyzer probe and preventing the analysis of blood. This can result in difficulties aspirating the sample or the plunger may get pushed out of the syringe. Smiths Medical has received no reports of injury associated with this issue.