Health Products

Sterrad Sterilizers (2013-11-07)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All serial numbers
Manufacturer
Advanced Sterilization Products
a Division of Ethicon
33 Technology Drive
Irvine
92618
California
UNITED STATES
D. Sterrad 100NX Sterilizer10104
All serial numbers
Manufacturer
Advanced Sterilization Products
a Division of Ethicon
33 Technology Drive
Irvine
92618
California
UNITED STATES
B. Sterrad 50 Sterilizer10050
All serial numbers
Manufacturer
Advanced Sterilization Products
a Division of Ethicon
33 Technology Drive
Irvine
92618
California
UNITED STATES
C. Sterrad NX Sterilizer10033
All serial numbers
Manufacturer
Advanced Sterilization Products
a Division of Ethicon
33 Technology Drive
Irvine
92618
California
UNITED STATES
A. Sterrad 100 S Sterilizer10101

Introduction

Recalled products
  1. Sterrad 100 S Sterilizer
  2. Sterrad 50 Sterilizer
  3. Sterrad NX Sterilizer
  4. Sterrad 100NX Sterilizer

Reason

ASP is notifying users to discontinue processing certain Karl Storz model flexible bronchoscopes and intubation fiberscopes in Sterrad system sterilizers. These endoscopes may contain a polyurethane material in the suction channel location that is not listed in the current claims for interior channels (Lumens) for processing in sterrad systems.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime