Immage Immunochemistry System (2013-08-19) Aug 19, 2013 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type III
Lot or serial number Brand name Companies Model or catalog number All lots B. Immage 800 Immunochemistry System Manufacturer Beckman Coulter Inc. 250 S. Kraemer Blvd. Brea 92821 California UNITED STATES A15445 All lots C. Reconditioned Immage Manufacturer Beckman Coulter Inc. 250 S. Kraemer Blvd. Brea 92821 California UNITED STATES 463804 All lots A. Immage Immunochemistry System Manufacturer Beckman Coulter Inc. 250 S. Kraemer Blvd. Brea 92821 California UNITED STATES 470000
Recalled products
Immage Immunochemistry System
Immage 800 Immunochemistry System
Reconditioned Immage
Beckman Coulter has verified that three fluid filters installed in the following locations on Immage or Immage 800 systems may prematurely fail (Clog): Reagent probe wash station, Sample probe wash station and Cuvette wash station.