In May 2013, Smiths Medical received a complaint concerning a Medfusion 4000 pump with thermal damage to the AC receptacle as a result of an electrical short. An in-depth investigation performed by a third-party laboratory, concluded that an electrical short can occur when saline solution or other conductive fluids (ie. Potassium chloride) enter the pump's AC power receptacle. When this happens, sparking and/or flames, accompanied by smoke, may appear at the power receptacle. The pump and AC power cord are constructed of flame retardant materials, therefore, the sparks/flames extinguish within seconds.