Health Products

Camino Advanced Monitor (2013-11-27)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
MEA1301804
Manufacturer
Integra Neurosciences Ltd.
also trading as Spembly Medical
Newbury Road
Andover, Hampshire
SP10 4DR
UNITED KINGDOM
A. Camino Advanced MonitorCAM01

Introduction

Recalled products
  1. Camino Advanced Monitor 

Reason

Integra Lifesciences has identified that if the optional SYS/DIAS mode is used on Camino advanced monitors (CAM01) with serial numbers formatted as MExxxxx04, the displayed values for the systolic and diastolic intracranial pressure (ICP) are reversed. But, if the CAM01 is used with an auxiliary bedside monitor, the bedside monitor displays these values correctly.