Dimension Vista Systems Aliquot Plates (2013-10-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
N2809513A. Dimension Vista Systems Aliquot Plates
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
10445210

Introduction

Recalled products
  1. Dimension Vista Systems Aliquot Plates 

Reason

Manufacturer has confirmed intermittent Aliquot plate barcode failure to read errors causing the plate to be discarded. The error is generated when the Aliquot plate is being loaded onboard the instrument or recalled from the Aliquot elevator for an add-on test. If the Aliquot plate cannot be read, it is discarded into the waste. Not all plates within a lot exhibit the failure. Not all instruments experience failures with the same lot of Aliquot plates.