Discovery PET/CT 710 System (2013-11-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
GE Medical Systems LLC
3200 North Grandview Blvd.
Waukesha
53188
Wisconsin
UNITED STATES
A. Discovery PET/CT 710 System5856030-X

Introduction

Recalled products
  1. Discovery PET/CT 710 System 

Reason

A software issue exists when using list mode replay, such that subsequent reconstructions with that data could contribute to an image with scatter over-estimation and over-correction. Reconstructed images with this issue will have many fewer counts than the acquired scan data and a high scatter fraction. This issue may affect studies that use list mode replay and are reconstructed after subsequent acquisitions. This defect may lead to misinterpretation of the image and/or inaccurate treatment decisions.