Investigations have determined that this lot has an increased frequency of Troponin I results greater than 0.02 ng/mL for samples which are less than 0.02 ng/mL upon additional testing. Any questions about previously reported results, the recommendation would be to consult with your resident clinical expert or physician in the setting where the test was performed.