Dimension Systems Reagent Cartridges (2013-11-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
FA4098
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
B. Dimension Xpand System - PBNP Flex Reagent CartridgeRF423A
FA4098
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
A. Dimension Chemistry System - PBNP Flex Reagent CartridgeRF423A

Introduction

Recalled products
  1. Dimension Chemistry System - PBNP Flex Reagent Cartridge
  2. Dimension Xpand System - PBNP Flex Reagent Cartridge

Reason

Siemens is advising customers of a change in QC values for the Thermo Scientific mas Cardioimmune XL control lot 1601 level 1. The manufacturer of the controls notified siemens of changes to the control product after the QC values were published, and as a consequence, the values provided with the Dimension N-Terminal Pro-Brain Natriuretic Peptide flex reagent lots GA4210, FC4168, FA4098 and FA4224 require revision.