ZYM B Reagent and API Biomedical Tests (2013-11-12) Nov 12, 2013 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number More than 10 numbers, contact manufacturer. Manufacturer Biomerieux SA chemin de l'Orme Marcy l'Etoile 69280 FRANCE B. API NH 10400 More than 10 numbers, contact manufacturer. Manufacturer Biomerieux SA chemin de l'Orme Marcy l'Etoile 69280 FRANCE C. API Listeria 10300 1001951820 Manufacturer Biomerieux SA chemin de l'Orme Marcy l'Etoile 69280 FRANCE A. ZYM B Reagent 70493
Recalled products
ZYM B Reagent
API NH
API Listeria
Investigations confirmed an activity issue on Zym B reagent whatever the color of Zym B ampule. This activity issue lead to a false negative result of some API biochemical tests using the Zym B reagent.