Health Products

Discovery PET/CT 600 and 690 Systems - Main Units (2013-12-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Hospitals, Healthcare Professionals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
GE Medical Systems LLC
3200 North Grandview Blvd.
Waukesha
53188
Wisconsin
UNITED STATES
B. Discover Pet/Ct 690 System - Main UnitS9164JC
Not applicable
Manufacturer
GE Medical Systems LLC
3200 North Grandview Blvd.
Waukesha
53188
Wisconsin
UNITED STATES
A. Discovery PET/CT 600 System - Main UnitS9116KC

Introduction

Recalled products
  1. Discovery PET/CT 600 System - Main Unit
  2. Discover Pet/Ct 690 System - Main Unit

Reason

A software issue exists when using list mode replay, such that subsequent reconstructions with that data could contribute to an image with scatter over-estimation and overcorrection. Reconstructed images with this issue will have many fewer counts than the acquired scan data and a high scatter fraction. This issue may affect studies that use list mode replay and are reconstructed after subsequent acquisitions.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime