Puritan Bennett Ventilator 840 (2013-12-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
Covidien LLC
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES
A. Puritan Bennett Ventilator 8404-840120DICU-EN

Introduction

Recalled products
  1. Puritan Bennett Ventilator 840

Reason

Covidien is taking voluntary corrective action in the field involving the Puritan Bennett 840 systems equipped with certain software revisions in response to reports by clients of diagnostic code XB0069, which results in an interruption in mechanical ventilation, owing to a software error. The frequency of occurrence of this code is extremely low.