Health Products

Veris 8600 Vital Signs Monitor - Systems (2013-11-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
300077
Manufacturer
Medrad Inc.
One Medrad Drive,
Indianola
15051-0780
Pennsylvania
UNITED STATES
F. Veris 8600 Vital Signs Monitor - Anesthesia with Temp. Wireless3011996W
37012
Manufacturer
Medrad Inc.
One Medrad Drive,
Indianola
15051-0780
Pennsylvania
UNITED STATES
G. Veris 8600 Vital Signs Monitor - Base Plus with Wireless3011992W
36432
Manufacturer
Medrad Inc.
One Medrad Drive,
Indianola
15051-0780
Pennsylvania
UNITED STATES
D. Veris 8600 Vital Signs Monitor - Capnography System3014268
37221
Manufacturer
Medrad Inc.
One Medrad Drive,
Indianola
15051-0780
Pennsylvania
UNITED STATES
E. Veris 8600 Vital Signs Monitor - Anesthesia with Wireless3011994W
21470
Manufacturer
Medrad Inc.
One Medrad Drive,
Indianola
15051-0780
Pennsylvania
UNITED STATES
B. Veris 8600 Vital Signs Monitor - Anesthesia System3011994
32902
Manufacturer
Medrad Inc.
One Medrad Drive,
Indianola
15051-0780
Pennsylvania
UNITED STATES
C. Veris 8600 Vital Signs Monitor - Cardiac System3011993
25655
Manufacturer
Medrad Inc.
One Medrad Drive,
Indianola
15051-0780
Pennsylvania
UNITED STATES
A. Veris 8600 Vital Signs Monitor - Base System3011992

Introduction

Recalled products
  1. Veris 8600 Vital Signs Monitor - Base System
  2. Veris 8600 Vital Signs Monitor - Anesthesia System
  3. Veris 8600 Vital Signs Monitor - Cardiac System
  4. Veris 8600 Vital Signs Monitor - Capnography Syst
  5. Veris 8600 Vital Signs Monitor - Anesthesia with Wireless
  6. Veris 8600 Vital Signs Monitor - Anesthesia with Temp. Wireless
  7. Veris 8600 Vital Signs Monitor - Base Plus with Wireless

Reason

The main board, part number 301641, installed in some Medrad Veris MR monitor units may be faulty and could lead to unexpected shutdown of the system while in use.  This would result in the loss of vital signs information from the monitor.  Failure of the monitor may not only result in an inability of the user to view vital signs information but this failure may also go unrecognized.  This may present an increased risk to patients who require that their vital signs be monitored while undergoing an MRI (Magnetic Resonance Image) exam such as those who are sedated, anesthetized or who are on critical medications.  Thus, it is required to continually monitor the system to ensure proper functioning.