Giraffe and Panda Warmers and Giraffe T-Piece Infant Resuscitation System (2013-11-19)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturerB. Panda Warmer
Manufacturer
Ohmeda Medical, A Division of Datex Ohmeda Inc.
8880 Gorman Road
Laurel
20723
Maryland
UNITED STATES
M1112198
More than 10 numbers, contact manufacturerC. Giraffe T-Piece Infant Resuscitation System
Manufacturer
Ohmeda Medical, A Division of Datex Ohmeda Inc.
8880 Gorman Road
Laurel
20723
Maryland
UNITED STATES
M1241420
More than 10 numbers, contact manufacturerA. Giraffe Warmer
Manufacturer
Ohmeda Medical, A Division of Datex Ohmeda Inc.
8880 Gorman Road
Laurel
20723
Maryland
UNITED STATES
M1118179

Introduction

Recalled Products
  1. Giraffe Warmer
  2. Panda Warmer
  3. Giraffe T-Piece Infant Resuscitation System

Reason

GE Healthcare has recently become aware of a potential safety issue wherein the oxygen and air wall or tank inlet fittings on the back panel, were reversed during assembly.  As a result, this may potentially have reversed air/oxygen gas concentrations, for example, a setting of 100% oxygen could have an output of 21% oxygen, and vice versa. The settings of the blender knob will no longer be accurate.