Health Products

R100 Ventilator System (2013-11-15)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Product Safety
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not Applicable
Manufacturer
Metran Co., Ltd.
2-12-18, Kawaguchi
Kawaguchi-Shi
332-0015
JAPAN
R100 Ventilator System10010410001

Introduction

Recalled Products

R100 Ventilator System

Reason

One report received of the actuation of ventilator alarm and buzzing sound after 16 hrs of patient oscillation in a hospital in Germany. The reason was a failure in the bonding of the diaphragm chamber producing leakage.