Platelet Concentrate Infusion Set and Lifeshield Plum and Lifeshield Plum Y-Type Blood Sets (2013-11-26) Nov 26, 2013 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number More than 10 numbers, contact manufacturer Manufacturer Hospira Inc. 275 North Field Drive Lake Forest 60045 Illinois UNITED STATES B. Lifeshield Plum Y-Type Blood Set 14212001 053835H Manufacturer Hospira Inc. 275 North Field Drive Lake Forest 60045 Illinois UNITED STATES C. Lifeshield Plum Blood Set 14210001 112274w Manufacturer Hospira Inc. 275 North Field Drive Lake Forest 60045 Illinois UNITED STATES A. Platelet Concentrate Infusion Set 14237001
Recalled Products
Platelet Concentrate Infusion Set
Lifeshield Plum Y-Type Blood Set
Lifeshield Plum Blood Set
There is a potential for certain blood sets to puncture the outer wall of non-iso compliant blood bags. The ISO piercing pin tip is sharper and longer than the port on the blood bag, which is very short in certain non-ISO blood bags and results in the leakage from the bag.