Kodak Directview DR 7500 System and DRX-Evolution Systems (2013-12-02) Dec 2, 2013 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Brand name Companies Model or catalog number 4236 F. DRX-Evolution System with Motorization-Fixed Wall Detector / Mobile CSI Table Detector Manufacturer Carestream Health Inc. 150 Verona Street Rochester 14608 New York UNITED STATES 1016138 4723 D. DRX-Evolution No System with Non-Motorized OTC Manufacturer Carestream Health Inc. 150 Verona Street Rochester 14608 New York UNITED STATES 1731645 4199 E. DRX-Evolution / Motorized OTC-FF Wallstand Manufacturer Carestream Health Inc. 150 Verona Street Rochester 14608 New York UNITED STATES 1731678 4208 B. DRX-Evolution System / Premium OTC-FF Wallstand - No Table -1DRXGOS Manufacturer Carestream Health Inc. 150 Verona Street Rochester 14608 New York UNITED STATES 1001700 4031 C. DRX-Evolution System / Premium OTC-FF Wallstand - Premium Table - 1DRXGOS Manufacturer Carestream Health Inc. 150 Verona Street Rochester 14608 New York UNITED STATES 1268457 75000150 A. Kodak Directview DR 7500 System Manufacturer Carestream Health Inc. 150 Verona Street Rochester 14608 New York UNITED STATES 879 1345
Recalled Products
Kodak Directview DR 7500 System
DRX-Evolution System / Premium OTC-FF Wallstand - No Table -1DRXGOS
DRX-Evolution System / Premium OTC-FF Wallstand - Premium Table - 1DRXGOS
DRX-Evolution No System with Non-Motorized OTC
DRX-Evolution / Motorized OTC-FF Wallstand
DRX-Evolution System with Motorization-Fixed Wall Detector / Mobile CSI Table Detector
Carestream has identified that the wall stand (WS) bucky may continue moving downward when releasing the motorized WS keypad button. The number of reported occurrences represents less than 1% of the DRX-Evolution systems installed since 2009. There have not been any reported occurrences for the DR 7500 system.