Health Products

Dimension Vista System - Cyclosporine (CSA Assay) (2013-12-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
12300BB
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
Dimension Vista System - Cyclosporine (CSA Assay) K4089

Introduction

Recalled Products

Dimension Vista System - Cyclosporine (CSA Assay) 

Reason

Siemens Helathcare Diagnostics has confirmed the possibility of variability in recovery of QC and patient results with the Dimension Vista CSA flex reagent cartridge lots 12300BB, 12318BB, and 13011BB. Internal studies have shown an upward trend in recovery over the three day open well est claim.