Health Products

Instrument Manager (2013-12-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
V8.12.01
Manufacturer
Data Innovations LLC
120 Kimball Ave. Suite 100
South Burlington
05403
Vermont
UNITED STATES
Instrument ManagerV8.12.01

Introduction

Recalled Products

Instrument Manager 

Reason

Subsequent child rules may not be evaluated if a rule attempts to divide by zero or another error occurs in a large rule set.  If rules are not evaluated it could produce undesired and/or incorrect results. This could cause harm to the patient if a medical decision is made based upon the erroneous result of the rule processed, or not processed.