Health Products

Somatom Definition Systems (2013-12-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
D. Somatom Definition Flash System - Main Unit10590000
All lots
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
B. Somatom Definition AS - Main Unit8098027
All lots
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
C. Somatom Definition Edge System - CT System10430603
All lots
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
A. Somatom Definition - Main Unit77 40 769

Introduction

Recalled Products
  1. Somatom Definition - Main Unit
  2. Somatom Definition AS - Main Unit
  3. Somatom Definition Edge System - CT System
  4. Somatom Definition Flash System - Main Unit

Reason

Somatom Definition CT systems running software version Syngo CT 2012B (VA44) may experience a malfunction which may lead to a system display freeze during interventional procedures.  When the function "auto save last displayed images as key images" is activated during an interventional procedure, the software configuration allows conflicting access to the CT system image database which may cause a system display freeze to occur. When this happens, normal interaction with the system is no longer possible. No error message is issued and only a system restart resolves the problem.  In case of a sudden system display freeze, the interventionalist will potentially not immediately recognize the malfunction and continue the interventional procedure. In view of the misleading information shown on the display, an injury of the patient cannot be fully ruled out.