Health Products

Oncor Expression, Oncor Impression and Artiste MV Linear Accelerators (2013-12-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
5414B. Artiste MV - Linear Accelerator - Main Unit
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
8139789
3779C.Oncor Impression IMRT Plus
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
5857912
5037A. Oncor Expression Digital Linear Accelerator
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
7360717

Introduction

Recalled products
  1. Oncor Expression Digital Linear Accelerator
  2. Artiste MV - Linear Accelerator - Main Unit
  3. Oncor Impression IMRT Plus

Reason

Siemens has become aware that it is possible for patients or operators to have their fingers trapped and pinched between the table top and the bottom frame of the 550 TXT treatment table, which is part of the affected systems listed in section 64(A). Secondly and unrelated, Siemens received a report on an incident in which a cable inside of the 550 TXT treatment table had been damaged due to contacting a moving part. The resulting short-circuit caused overheating of an electronic circuit board inside the table base.