Health Products

Hemochron Jr. Citrate APTT Test Cuvette (2013-12-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
E3JCC014-P2
Manufacturer
International Technidyne Corp.
8 Olsen Avenue
Edison
08820
New Jersey
UNITED STATES
A. Hemochron Jr. Citrate APTT Test Cuvette J103C

Introduction

Recalled products
  1. Hemochron Jr. Citrate APTT Test Cuvette 

Reason

The manufacturer ITC has determined that some citrate activated partial thromboplastin time (APTT) cuvettes may recover higher than expected results in normal individuals. Investigation into the affected lot identified that when performing tests on non-heparinized, e.g., normal blood samples, results may average higher than historic values. Test results performed on patients receiving heparin therapy are not affected.