Advia 1650 Chemistry System Analyzer and Advia 1200 Chemistry System Instrument (2013-12-19)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
CA12240082
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
B. Advia 1200 Chemistry System Instrument02475632
CA175542364
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
A. Advia 1650 Chemistry System Analyzer06363413

Introduction

Recalled products
  1. Advia 1650 Chemistry System Analyzer
  2. Advia 1200 Chemistry System Instrument

Reason

In the Advia 1200 chemistry system (Software versions V2.00 and V2.01) and Advia 1650 chemistry system (Software versions V3.52 and V4.01), an overflow flag "/////" is displayed in the result column for a test when the algorithm for that test cannot calculate a result or when the concentration of a test cannot be determined because it is outside the absorbance limits. "/////" is not a usable or reportable result however, it is possible for the user to misinterpret or edit this result.