A. ABX Pentra 400 Clinical Chemistry System - Controls
A11A01653
Introduction
Recalled products
ABX Pentra 400 Clinical Chemistry System - Controls
Reason
During a complaint investigation at the manufacturer (Horiba ABX SAS), a shift in the target value of the ABX Pentra N control was detected for Potassium-E values related to the ABX Pentra 400 only. Clinical laboratories will observe the Potassium N control result will be reported in the lower range of the N control target range. The original target value annex shipped with the ABX Pentra N control lot: 1300801 identified the potassium target value as 4.00 mmol/L with a target range of 3.70 - 4.30 mmol/l (equivalent to 15.6 mg/dL with a target range of 14.4 - 16.8 mg/dL).
Horiba ABX SAS (the manufacturer) has reassigned the target value and range and updated the annex for this lot number to reflect a target value of 3.76 mmol/L and a target range of 3.46 - 4.06 mmol/L (equivalent to 14.7 mg/dL with a target range of 13.5 - 15-.9 mg/dL) for potassium. This issue is only applicable to Potassium-E values reported on the ABX Pentra 400 chemistry analyzer with the ISE module only. No other analyzers or other analyte values are affected.