Brightview Gamma, XCT, and X Camera Systems (2013-12-24)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Philips Medical Systems (Cleveland), Inc.
595 Miner Road
Cleveland
44143
Ohio
UNITED STATES
D. Brightview X System with 3/8 inch Crystal (Upgradeable Version)453560829241
All lots
Manufacturer
Philips Medical Systems (Cleveland), Inc.
595 Miner Road
Cleveland
44143
Ohio
UNITED STATES
B. Brightview Gamma Camera System with 3/8 inch Crystal and Detector Caudal Cephalic Tilt2170-3002A
All lots
Manufacturer
Philips Medical Systems (Cleveland), Inc.
595 Miner Road
Cleveland
44143
Ohio
UNITED STATES
C. Brightview XCT Imaging System - Tilt System453560462131
All lots
Manufacturer
Philips Medical Systems (Cleveland), Inc.
595 Miner Road
Cleveland
44143
Ohio
UNITED STATES
A. Brightview Gamma Camera System with 3/8 inch Crystal2170-3000A

Introduction

Recalled products
  1. Brightview Gamma Camera System with 3/8 inch Crystal
  2. Brightview Gamma Camera System with 3/8 inch Crystal and Detector Caudal Cephalic Tilt
  3. Brightview XCT Imaging System - Tilt System
  4. Brightview X System with 3/8 inch Crystal (Upgradeable Version)

Reason

Philips has become aware of several software issues which will be corrected with the release of a new software version.

Problem 1: When positioning for a relative 180 degree non circular acquisition the second (bottom) detector was not positioned correctly by an operator, causing an inappropriate rotation orbit calculation for the patient being imaged. When the acquisition was started and the detectors began to move into position, the incorrect calculation resulted in the detectors coming in contact with the patient.

Problem 2: On separate occasions, as the gantry system was moving to a relative 90 scan position of the detectors, a patient or operator moved their hands or arms into the path of the detectors in the area of where the detectors meet resulting in the detectors coming in contact with the patient and operator.

Problem 3: While a user was manually positioning the patient for their exam and translating the pallet into the gantry bore, the pallet catcher that supports the imaging table was higher than the imaging pallet resulting in a pinch point or table tilt if not engaged with the catcher. This problem can only occur on the Brightview X or the Brightview XCT system.