System 1 Express Quick Connect (QFC1716INT) (2013-12-18)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
5984372
Manufacturer
Steris Corporation
5960 Heisley Road
Mentor
44060
Ohio
UNITED STATES
A. System 1 Express Quick Connect (QFC1716INT)QXCXXXXINT

Introduction

Recalled products
  1. System 1 Express Quick Connect (QFC1716INT)

Reason

QFC1716INT is an accessory to the System 1 Express sterile processing system. This quick connect consists of labeling and a flow unit made up of tubing, tethers and adapters used for connecting ports of specified Fujinon devices to flexible processing trays/containers placed within the System 1 Express. Steris has learned that the suction connector barb adapter on the QFC1716INT does not properly fit the suction connector port of the Fujinon 530 series endoscopes, and consequently disconnects from the endoscope prior to reprocessing.