Health Products

Dimension Reaction Vessels (2013-12-19)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lotsA. Dimension Reaction Vessels 
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
RXV1A

Introduction

Recalled products
  1. Dimension Reaction Vessels 

Reason

Siemens Healthcare Diagnostic has observed an increase in the frequency of complaints for vessel feeder errors. The errors observed include "670: vessel feeder track sensor stuck on full", and "749: vessel track empty". Complaints have been confirmed with several Dimension reaction vessel lots. A low percentage of reaction vessels from those lots have been observed to have flashing on the flange of the vessel that may jam in the vessel track and generate an error.