Brilliance CT Big Bore System - Main Unit (2013-12-19)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Philips Medical Systems (Cleveland), Inc.
595 Miner Road
Cleveland
44143
Ohio
UNITED STATES
A. Brilliance CT Big Bore System - Main Unit 4535 670 88051

Introduction

Recalled products
  1. Brilliance CT Big Bore System - Main Unit 

Reason

Philips has become aware that the Brilliance CT Big Bore was found to be out of tolerance for radio frequency emissions. Specifically, at the 48mhz frequency, the testing indicated the Brilliance CT Big Bore was 3.5 dBuV/meter higher than the applicable International Electronic Commission (IEC) 60601-1-2 standard specification.