Resuscitation Units - Giraffe Warmer, Panada Warmer and T-Piece (2013-12-13) Dec 13, 2013 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Brand name Companies Model or catalog number More than 10 numbers, contact manufacturer. B. Panda Warmer Manufacturer Ohmeda Medical, A Division of Datex Ohmeda Inc. 8880 Gorman Road Laurel 20723 Maryland UNITED STATES M1112198 More than 10 numbers, contact manufacturer. C. T-Piece Resuscitation Unit Manufacturer Ohmeda Medical, A Division of Datex Ohmeda Inc. 8880 Gorman Road Laurel 20723 Maryland UNITED STATES M1223628 More than 10 numbers, contact manufacturer. A. Giraffe Warmer Manufacturer Ohmeda Medical, A Division of Datex Ohmeda Inc. 8880 Gorman Road Laurel 20723 Maryland UNITED STATES M1118179
Recalled Products
Giraffe Warmer
Panada Warmer
T-Piece Resuscitation Unit
GE Healthcare has recently become aware of a potential safety issue associated with the loosening of the blender knob shaft on all resuscitation units. A loose blender knob shaft may potentially affect the accuracy of the blended output of Air/O2