Health Products

Tracoe Twist Tracheostomy and Laryngectomy Tubes (2013-07-19)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
Not ApplicableH.    Tracoe Twist Tracheostomy Tube with Oxygen Supply Line
Manufacturer
Tracoe Medical GmbH
Reichelsheimer Str. 1/3
Niederholm
55268
GERMANY
308-04 to 308-10
Not ApplicableI.      Tracoe Twist Tracheostomy Tube with Low Pressure Cuff and Air Supply Line for Speaking
Manufacturer
Tracoe Medical GmbH
Reichelsheimer Str. 1/3
Niederholm
55268
GERMANY
309-04 to 309-10
Not ApplicableF.    Tracoe Twist Tracheostomy Tube with Low Pressure Cuff and Suction Line
Manufacturer
Tracoe Medical GmbH
Reichelsheimer Str. 1/3
Niederholm
55268
GERMANY
306-04 to 306-10
Not ApplicableG.    Tracoe Twist Tracheostomy Tube with Low Pressure Cuff and Oxygen Supply Line
Manufacturer
Tracoe Medical GmbH
Reichelsheimer Str. 1/3
Niederholm
55268
GERMANY
307-04 to 307-10
Not ApplicableD.    Tracoe Twist Fenestrated Tracheostomy Tube with two Inner Cannulas
Manufacturer
Tracoe Medical GmbH
Reichelsheimer Str. 1/3
Niederholm
55268
GERMANY
304-04 to 304-10
Not ApplicableE.    Tracoe Twist Laryngectomy Tube with two Inner Cannulas
Manufacturer
Tracoe Medical GmbH
Reichelsheimer Str. 1/3
Niederholm
55268
GERMANY
305-05 to 305-10
Not ApplicableB.    Tracoe Twist Fenestrated Tracheostomy Tube with Low Pressure Cuff
Manufacturer
Tracoe Medical GmbH
Reichelsheimer Str. 1/3
Niederholm
55268
GERMANY
302-04 to 302-10
Not ApplicableC.    Tracoe Twist Tracheostomy Tube with two Inner Cannulas
Manufacturer
Tracoe Medical GmbH
Reichelsheimer Str. 1/3
Niederholm
55268
GERMANY
303-04 to 303-10
Not ApplicableA.    Tracoe Twist Tracheostomy Tube with Low Pressure Cuff
Manufacturer
Tracoe Medical GmbH
Reichelsheimer Str. 1/3
Niederholm
55268
GERMANY
301-04 to 301-10

Introduction

Recalled Products

A. Tracoe Twist Tracheostomy Tube with Low Pressure Cuff
B. Tracoe Twist Fenestrated Tracheostomy Tube with Low Pressure Cuff
C. Tracoe Twist Tracheostomy Tube with two Inner Cannulas
D. Tracoe Twist Fenestrated Tracheostomy Tube with two Inner Cannulas
E. Tracoe Twist Laryngectomy Tube with two Inner Cannulas
F. Tracoe Twist Tracheostomy Tube with Low Pressure Cuff and Suction Line
G. Tracoe Twist Tracheostomy Tube with Low Pressure Cuff and Oxygen Supply Line
H. Tracoe Twist Tracheostomy Tube with Oxygen Supply Line
I. Tracoe Twist Tracheostomy Tube with Low Pressure Cuff and Air Supply Line for Speaking

Reason

In some case the inner cannula locking system of the tracoe twist tracheostomy tube maybe too tight thus causing difficulties for the user in changing of the inner cannula.
Get told about the next health product recall.
Get health product recall alerts