A. Immulite 2000 System - Androstenedione Assay
B. Immulite System - Androstenedione Assay
C. Immulite 1000 System - Androstenedione Assay
Reason
Siemens Healthcare Diagnostics has confirmed over-recovery of samples with concentrations greater than 5.5 ng/mL. Siemens has determined that these lots are linear up to a concentration of 5.5 ng/mL, instead of 10 ng/mL as stated in the instructions for use (IFU). Quality control materials at concentrations between 5.5 and 10 ng/mL will detect this issue.