Immulite Systems - Androstenedione Assay (2013-12-24)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
409 and aboveB. Immulite System - Androstenedione Assay
Manufacturer
Siemens Healthcare Diagnostics Products Ltd.
Glyn Rhonwy
Llanberis, Gwynedd
LL55 4EL
UNITED KINGDOM
LKAO1
409 and above C. Immulite 1000 System - Androstenedione Assay
Manufacturer
Siemens Healthcare Diagnostics Products Ltd.
Glyn Rhonwy
Llanberis, Gwynedd
LL55 4EL
UNITED KINGDOM
LKAO1
314 and aboveA.  Immulite 2000 System - Androstenedione Assay
Manufacturer
Siemens Healthcare Diagnostics Products Ltd.
Glyn Rhonwy
Llanberis, Gwynedd
LL55 4EL
UNITED KINGDOM
L2KAO2

Introduction

Recalled Products

A.  Immulite 2000 System - Androstenedione Assay
B. Immulite System - Androstenedione Assay
C. Immulite 1000 System - Androstenedione Assay

Reason

Siemens Healthcare Diagnostics has confirmed over-recovery of samples with concentrations greater than 5.5 ng/mL. Siemens has determined that these lots are linear up to a concentration of 5.5 ng/mL, instead of 10 ng/mL as stated in the instructions for use (IFU). Quality control materials at concentrations between 5.5 and 10 ng/mL will detect this issue.