Health Products

Advia Centaur, Centaur CP, and Centaur XP (2013-12-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
B. Advia Centaur04610049
All lots
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
C. Advia Centaur CP04610049
All lots
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
A. Advia Centaur XP04610049

Introduction

Recalled Products

A. Advia Centaur XP
B. Advia Centaur
C. Advia Centaur CP

Reason

Siemens has received and confirmed complaints of elevated cortisol calibration slopes that are outside of the defined reagent lot specific calibration slopes when using cortisol reagents prior to the end of the reagent shelf life. When this occurs, the system will flag the failed calibration of the assay, and quality control and patient results can not be generated.