Eightcheck 3WP X-TRA Control (2013-05-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
3120
Manufacturer
Sysmex Corporation
1-5-1 Wakinohama-kaigandori
Chuo-ku, Kobe, Hyogo-ken
651-0073
JAPAN
A. Eightcheck 3WP X-TRA Control14030040

Introduction

Recalled Products

Eightcheck 3WP X-TRA Control

Reason

Control product Eightcheck 3WP X-TRA lot 3120, had an error on the Poch-100I assay barcode sheet for mean platelet volume (MPV), level 3 (l3). The barcode sheet had an incorrect MPV target value of 14.2 (correct value: 10.6). However, the table of values on the assay sheet was correct with a value of 10.6 for MPV l3. All other assay values for lot 3120 were correct. The barcode discrepancy was reported to the technical assistance center which led to discovery of incorrrect bar code value for MPV- l3 assay target. Customer was preparing the Poch-100Ito use the new Eightcheck 3WP X-TRA lot 3120 and noticed the discrepancy. The SAI assay confimed the discrepancy and initiate corrective action.