Health Products

Xio Radiation Treatment Planning System (2013-12-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
XiO Release 4.50.00  
Manufacturer
Elekta Business Area Software Systems Impac Medical Systems Inc.
13723 Riverport Drive, Suite 100
Maryland Heights
63043
Missouri
UNITED STATES
A. Xio Radiation Treatment Planning System 9200

Introduction

Recalled products
  1. Xio Radiation Treatment Planning System 

Reason

When the bolus is present, the effective depth (with bolus) to the weight point returned on the source data report is incorrect. The problem occurs when the bolus polygon incorrectly invades the patient contour and overlaps an area where the CT density is much different than bolus density. This is more likely to occur with convex patient surfaces.