Health Products

Vitek 2, Vitek 2 Compact 30, and Vitek 2 Compact 60 system susceptibility test cards (2013-12-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lotsB. Vitek 2 Compact 60 System - Susceptibility Test Cards
Manufacturer
Biomerieux Inc.
595 Anglum Road
Hazelwood
84894
Missouri
UNITED STATES
410101
All lotsC.    Vitek 2 System - Susceptibility Test Cards
Manufacturer
Biomerieux Inc.
595 Anglum Road
Hazelwood
84894
Missouri
UNITED STATES
410101
All lotsA. Vitek 2 Compact 30 System - Susceptibility Test Cards
Manufacturer
Biomerieux Inc.
595 Anglum Road
Hazelwood
84894
Missouri
UNITED STATES
410101

Introduction

Recalled Products
  1. Vitek 2 Compact 30 System - Susceptibility Test Cards
  2. Vitek 2 Compact 60 System - Susceptibility Test Cards
  3. Vitek 2 System - Susceptibility Test Cards

Reason

There have been recent complaints that the Vitek 2 Meropenem (Mem, Mm02N) is overcalling resistance for isolates of Aeromonas hydrophila. False resistance was confirmed when isolates from customer sites were tested internally by broth microdilution (BDM) the reference method used for development of MEM02N.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime