Dimension Vista System - Enzyme Calibrator (2014-01-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
3FD034
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
A. Dimension Vista System - Enzyme Calibrator KC310

Introduction

Recalled products
  1. Dimension Vista System - Enzyme Calibrator 

Reason

There have been reports of quality control (QC) shifts with pseudocholinesterase (PCHE) and imprecision during PCHE calibration with Dimension Vista enzyme 1 calibrator (ENZ 1 cal) lots 3FD034 and 3HD012. Investigation by Siemens confirms a loss of PCHE activity in the affected lots of enzyme 1 calibrator, causing an elevation of qc by as much as 80% and in some instances, imprecision of replicates during calibration.