Health Products

Da Vinci Surgical System - IS3000 and Da Vinci SI System - Vision System Cart (2013-10-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
273456B. Da Vinci SI System - Vision System Cart
Manufacturer
Intuitive Surgical
1266 Kifer Road, Building 101
Sunnyvale
94086
California
UNITED STATES
VS3000
SH1104A. Da Vinci Surgical System - IS3000
Manufacturer
Intuitive Surgical
1266 Kifer Road, Building 101
Sunnyvale
94086
California
UNITED STATES
IS3000

Introduction

Recalled Products

Da Vinci Surgical System - IS3000 and Da Vinci SI System - Vision System Cart

Reason

It has been identified that the metalized coating on the IS3000 HD lamp module (Part number: 950093-05) may be incompatible with the illuminator (Part number: 951183-06) control board, leading to an erroneous red preventive maintenance advisory (PMA) that shows up on all monitors and cannot be dismissed by the user.  This error does not affect the lighting of the lamp module in the normal fashion, and there are no reports of lamp failure that have been associated with this issue.