Health Products

Optima X-Ray Systems (2013-12-05)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
Not applicableB. Optima XR200AMX X-Ray System - Digital
Manufacturer
GE Medical Systems, LLC
3000 North Grandview Blvd.
Waukesha
53188
Alabama
UNITED STATES
5421698
Not applicableA. Optima XR220AMX X-Ray
Manufacturer
GE Medical Systems, LLC
3000 North Grandview Blvd.
Waukesha
53188
Alabama
UNITED STATES
5555000-5

Introduction

Recalled Products

Optima  X-Ray Systems 

Reason

Intermittently when a large number of patient exams are retrieved from the HIS/RIS system, the exams when displayed on the mobile X-ray system are not properly displayed. If the user continues to attempt to select one of the exams when this issue has occurred, the system will update the screen and in doing so, potentially select a different exam/patient.