Health Products

TDHisto/Cyto Software (2013-12-09)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
V. V11.71.A and above 
Manufacturer
Technidata S.A.S.
387, Avenue Jean Kuntzmann
Montbonnot Saint Martin
38330
FRANCE
TDHisto/Cyto Software TD-TDH-01

Introduction

Recalled Products

TDHisto/Cyto Software 

Reason

For TDHisto/Cyto software versions v11.71.a and higher, there is a risk of sending review requests with "valid" status after a certain sequence of screens (see the communication to users for details). The issue may lead to releasing patient information that has not been clinically validated. This problem only impacts sites using the "screen" validation mode.