Health Products

Vitros White Correction Factor Slides (2014-01-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
9052-0045-6358
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
Vitros White Correction Factor Slides J02315

Introduction

Recalled Products

Vitros White Correction Factor Slides 

Reason

OCD has initiated a recall as it has determined that one lot of Vitros White Correction Factor Slides was released with an incorrect assay value for the 460 nm wavelength filter. The labeling for the Vitros White Reference Slides lot 9052-0045-6358 incorrectly indicates an assay value of 0.3785 for the 460 nm filter versus the actual assay value of 0.8658. All other assay values listed are correct. The 460 nm wavelength is only used to determine the predicted concentration when processing Vitros TBIL slides or Vitros BUBC slides.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime