Carbide Cutting Burrs (2013-12-19) Dec 19, 2013 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number More than 10 numbers, contact manufacturer Manufacturer The Anspach Effort Inc. 4500 Riverside Drive Palm Beach Gardens 33410 Florida UNITED STATES B. Bur - Carbide Fluted Ball More than 10 numbers, contact manufacturer C073015750 Manufacturer The Anspach Effort Inc. 4500 Riverside Drive Palm Beach Gardens 33410 Florida UNITED STATES A. Bur - Carbide Fluted Ball, Standard S-2SB-C
Recalled Products
Carbide Cutting Burrs
The laser marking location is out of specification which causes it to align with the distal bearing of the attachment and can lead to a potential cutting Bur fracture.