Health Products

Neptune 2 Waste Management System - Rover (2014-01-24)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Stryker Instruments
4100 East Milham Ave.
Kalamazoo
49001
UNITED STATES
Neptune 2 Waste Management System - Rover 0702-001-000

Introduction

Recalled Products

Neptune 2 Waste Management System - Rover 

Reason

The purpose of this correction is to upgrade the Neptune 2 Waste Management System to the new approved specifications. Previously, Stryker had received reports that the Neptune 2 had been connected directly to a chest tube, postoperatively which resulted in fatalities and injuries. The new approved specifications include a change in technology, a control mechanism change as well as a modification to the canister vacuum level regulator (from mechanical to electrical). The Neptune 2 has been re-designed to address the hazards and harms outlined in previously recalls.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime