B. Advia Chemistry System - Advia 2400 Analyzer Class 2
01714935
All lots
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
A. Advia Chemistry System - Advia 1800 Analyzer Class 2
01412173
Introduction
Recalled Products
Advia Chemistry Systems
Reason
Siemens Healthcare Diagnostics is conducting a recall for Advia 1800 and Advia 2400 ISE firmware v2.23. Advia Chemistry Chloride Electrodes may fail prematurely during calibration due to a chloride bias parameter error. The operator will receive an ng flag which requires immediate action to be taken on the calibration samples.