Health Products

CPAP Boussignac (2014-01-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
171111EJ
Manufacturer
Vygon S.A.
5 Rue Adeline
Ecouen
95440
FRANCE
Appareils de VPPC Boussignac  5570.13

Introduction

Recalled Products

CPAP Boussignac 

Reason

Boussignac CPAP units from lots 171111EJ and 240112EK are being withdrawn from the market because of reports of the connection holding the oxygen tubing were broken as received (Out-of-box failure).
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime